High Performance Thickness Gauge
Features:
Setups: 25 manufacturer presets that can be edited by the user.
Gates: 2 fully functional gates with adjustable thresholds.
Windows® PC interface software: Bi-directional transfers of grid data, waveforms, and setups.
Full Feature Lockout:
Password locks or enables some of the advanced features. Locked: Use THOR as a digital thickness gauge. Unlocked: THOR is a fully functional A-Scan scope.
Data Logger (Internal)
1000 pages of memory (alpha numeric storage)—1 main celll reading, 1 waveform, 2 comment lines, and 16 sub
readings per page.
OBSTRUCT for indicating inaccessible locations.
Transducer
Transducer Types:
- Dual Element (1 to 10 MHz)
- Single Element Contact (500 KHz to 10 MHz)
- Single Element Delay Line ( 5 to 22 MHz )
Locking quick disconnect “00” LEMO connectors.
Standard 4 foot cable.
Custom transducers and lengths cables available for
special applications.
Power Source
Four 1.5V alkaline or 1.2V NiCad C cells
100—240 VAC input 12VDC power supply.
Typically operates for 30 hours on alkaline and 15 hours on NiCad (charger not included.)
Battery icons indicate remaining life.
Measuring Range:
Measures from 0.010 to 480
inches (0.63 to 12 Meters) in
steel. Switch to select english
or metric units.
Resolution:
Standard Mode—
.001 inches ( 0.01 millimeters )
High Resolution mode—
.0001 inches ( 0.001 mm )
Velocity Range:
.0500 to .3937 in/ms
1270 to 10,000 meters/sec
Selectable pulse repetition rate
allows for high speed scanning
or low power operation.
Single and Two point calibration
capability.
Certification
Factory calibration traceable to
national standards. |
Physical
Size:
Width (8.5in/216 mm)
Height (6.5 in/165 mm)
Depth (2.0 in/51 mm)
Weight:
2lbs. 10 ounces (1.2 Kg)
(with batteries).
Keyboard:
Membrane switch pad with
tactile keys.
Operating Temperature:
-4° to 122°F (-20° to 50°C)
Case:
Extruded aluminum body with
nickel-plated aluminum end
caps (gasket sealed).
Data Output:
RS232 port out Windows®
interface software.
Display:
Backlit LCD—3.5“W x 2.0”H
Ultrasonic Specifications
Measurement Modes:
Pulse-Echo (flaws, pits)
Echo-Echo (thru paint,
precision delay line)
Pulser:
Tone-burst (push-pull).
Selectable 1, 2, or 4 pulses.
Selectable frequencies (MHz):
0.5, 1, 2.25, 5, Hi Freq.
Receiver:
Low noise pre-amplifier.
Adjustable gain control.
Timing:
160 MHz 10 ppm TCXO
controlled digitizer.
Memory:
8 megabit non-volatile ram.
Display
Display Views:
- Rectified ( flaw view )
- RF ( full waveform view )
- Digital thickness readout
Multi-function 6 digit graphic
LCD, backlit for use in poor light conditions.
Measurements displayed in inches, inches/microsecond.
Millimeters and meters/second.
Bar graph indicates stability of
reading. |